Loading...
18
RETURN TO TABLE
OF CONTENTS
Foot Control
Function Pedals
NOTE
The HAND CONTROL must be
plugged into the table in order for the
foot control to function.
The foot control has 6 control
pedals; Trendelenburg,
Reverse Trendelenburg, Height
Up, Height Down, Lateral Tilt
Left, and Lateral Tilt Right.
Installation
Plug the cord of the Foot Con-
trol into the port labeled FOOT
CONTROL located on the bot-
tom, patient’s right side of the
table top
Priority
The Hand Control has priority
over the Foot Control. Should a
function be activated by the
Foot Control a person can over-
ride the Foot Control command through use of the Hand Control.
Display
The display on the hand control will display messages when using the Foot
Control.
WARNING
Should the table
malfunction, imme-
diately remove your foot from
the footswitch. For emergency
operation of the table refer to
Emergency Override Panel.
CA800700
OMI
7100
TRENDEL-
ENBURG
REVERSE
TRENDEL-
ENBURG
TABLE
UP
TABLE
DOWN
LATERAL
TILT
RIGHT
LATERAL
TILT
LEFT
Operation

Other manuals for Schaerer 7300