Foot Control
Function Pedals
NOTE
The HAND CONTROL must be
plugged into the table in order for the
foot control to function.
• The foot control has 6 control
pedals; Trendelenburg,
Reverse Trendelenburg, Height
Up, Height Down, Lateral Tilt
Left, and Lateral Tilt Right.
Installation
• Plug the cord of the Foot Con-
trol into the port labeled FOOT
CONTROL located on the bot-
tom, patient’s right side of the
table top
Priority
• The Hand Control has priority
over the Foot Control. Should a
function be activated by the
Foot Control a person can over-
ride the Foot Control command through use of the Hand Control.
Display
• The display on the hand control will display messages when using the Foot
Control.
WARNING
Should the table
malfunction, imme-
diately remove your foot from
the footswitch. For emergency
operation of the table refer to
Emergency Override Panel.