REFLEX
Indicates the table is being moved to a reflex position. (Reflex is not available in
the Reverse Position).
RETURN TO LEVEL
Informs the operator the entire table top is being moved to a level position.
RECHARGE BATTERY / PLUG CORD IN NOW
Informs the operator the battery charge has fallen below 1/6 of the capacity and
that it is imperative that the power cord be plugged in immediately to allow the
batteries to be recharged.
BATT LEVEL LOW / RECHARGE BATTERY
The battery charge is less than 1/3 of capacity.
BATT LEVEL MED
The battery charge is between 1/3 and 2/3 of capacity.
BATT LEVEL FULL
Battery charge is more than 2/3 of capacity.
CHARGING BATTERY
Table is connected to an electrical outlet and the batteries are being recharged.
CALIBRATING NEUTRAL POSITION
Informs the Service Technician that the “neutral position” is being calibrated and
the data is being entered into the main controller.
SERVICE REQUIRED
Informs the operator that service is required by a qualified service technician.
The self diagnostic function of the table has detected a malfunction.
ERROR CODE Exx
Provides information to a qualified service technician as to what type of mal-
function exist. The operator should write the error code down and relate it to the
service technician when the service call is required.