What to do if my Schaerer Medical Equipment table seems powerless?
- NNicholas AguilarAug 17, 2025
If your Schaerer Medical Equipment table appears to have no power, even when plugged in, there are several potential causes. First, check if the facility circuit breaker has tripped, and if so, determine the cause and reset or replace it. Second, inspect the 5 amp fuses in the power cord receptacle and replace any that are blown. Third, check the RFI filter and replace it if necessary. Fourth, replace the line power pilot lamp if it's burned out. Finally, examine the wires and connections between the power cord receptacle and transformer, cleaning dirty connections, tightening loose ones, and replacing any broken components.