SECTION IV
MAINTENANCE / SERVICE
SECTION IV
MAINTENANCE / SERVICE INSTRUCTIONS
CAUTION
ISO Viscosity Grade 32 Premium Hydrau-
lic Fluid is only type of oil which can be
used in hydraulic system of this table. Failure to do
so could result in damage to hydraulic system.
(7) If oil level is low, insert a funnel into oil check
hole and add ISO Viscosity Grade 32 Premium
Hydraulic Fluid until proper oil level, as de-
scribed in step 6, is reached.
(8) Screw breather cap (3) into oil check hole.
(9) Install hole plug (1) on base weldment (2).
Figure 4-1. Checking / Adding Oil
4.1 Introduction
DANGER
Refer to Operator Manual for
complete instructions on operating
surgery table. Failure to do so could result in
personal injury.
NOTE
Perform an operational test on surgery table after
repair is completed to confirm repair was properly
made and that
all
malfunctions were repaired.
The following paragraphs contain removal, installa-
tion, repair, and adjustment procedures for surgery
table.
4.2 Checking Oil Level / Adding Oil To
Reservoir
A. Checking / Adding Oil
(1) Lower TABLE DOWN function until column
assembly is fully retracted.
NOTE
Pressing LEVEL button on the hand control will
accomplish following step.
(2) Using hand control or emergency override panel,
move Trendelenburg, seat, and lateral tilt
functions until table top is level.
(3) Using the hand control or emergency override
panel, unlock the floor locks.
(4) Remove hole plug (1, Figure 4-1) from base
weldment (2).
(5) Unscrew breather cap (3) from oil check hole.
(6) Using a flashlight, check oil level. Oil should be
approximately 1 in. (2.54 cm) below top of oil
check hole opening, which is when oil just starts
to cover top of motor pump housing.
Page 4-1Printed in U.S.A.
CA7617
SMI
7100
OIL
2
OIL
CHECK
HOLE
1
3
1/16"
(1.6 mm)
1"
(2.54cm)
1
3
© Schaerer Mayfield USA, Inc. 2004