SECTION IV
MAINTENANCE / SERVICE
Page 4-39Printed in U.S.A.
4.24 Discharge Filter Replacement
A . Removal
DANGER
Table is very heavy. Use an assistant
to lay table onto its side. Failure to do
so could result in serious personal injury.
(1) Lay table onto its left side.
(2) Remove six screws (1, Figure 4-39) and main
cover (2) from base weldment (3).
(3) Remove two screws (4), two screws (5),
mounting bracket (6), and partially remove
control box (7) from base weldment (3).
(4) Remove two screws (8), two screws (9),
mounting bracket (10), two spacers (11), and
partially remove manifold assembly (12) from
base weldment (3).
(5) Remove fitting connector (13), fitting seal (14),
and tube assembly (15) from motor pump (16).
(6) Work tube assembly (15) around and pull tube
and discharge filter (17) out base weldment (3).
Figure 4-39. Discharge Filter Removal / Installation
1
2
3
4
5
7
6
9
11
8
10
12
16
13
14
15
17
18
CA817100
© Schaerer Mayfield USA, Inc. 2004