SECTION IV
MAINTENANCE / SERVICE
4.3 Calibration Of Return To Level /
Neutral Position
A. Calibration
(1) Move table to a flat, level surface; then press
ENABLE button on hand control to engage floor
locks.
(2) Using a protractor, check to make sure base of
table is level (± 1°) from front-to-rear and from
side-to-side. See Figure 4-2.
(3) If installed, remove head section from back
section.
(4) If installed, remove cushion from back section.
(5) Place protractor at Position A; then using hand
control, move LATERAL TILT LEFT and
LATERAL TILT RIGHT functions until back
section is level (± 0.5 °) in side-to-side direction.
Figure 4-2. Leveling Table Top
Page 4-2Printed in U.S.A.
CA805800
LATERAL
TILT LEFT
REVERSE
TRENDELENBURG
LATERAL
TILT RIGHT
NORMAL
POSITION
REVERSE
POSITION
TREND REV TREND
HEIGHT UP
HEIGHT DOWN
TILT L TILT R
SEAT UP SEAT DOWN
FLEX
REFLEX
ENABLE
DISABLE
LOCK
UNLOCKLEVEL
SMI
7300
NORMAL
POSITION
REVERSE
POSITION
TREND
REV TREND
HEIGHT UP
HEIGHT DOWN
TILT L
TILT R
SEAT UP SEAT DOWN
FLEX
REFLEX
ENABLE
DISABLE
LOCK
UNLOCKLEVEL
SMI
7300
SMI
7100
POSITION
"C"
POSITION
"B"
POSITION
"A"
PROTRACTOR
BACK
SECTION
TABLE
BASE
0°
(+ / - 1°)
0°
(+ / - 1°)
TRENDELENBURG