4.5.4 Erasing the data memory
You can erase all stored measurement datasets.
4.5.5 Downloading calibration data
You can download calibration data:
z to the display
via the memory menu
via the calibration menu
z to the interface (see page 39)
Download to display via
memory menu
While the calibration data is being displayed, you can:
1 Open the memory menu with <RCL>.
2 If necessary, scroll with <RCL> until Sto disp or Sto prt is
displayed.
3 Press <RCL_> to display the Clr All function.
4Press <OK> to erase the entire contents of the memory.
or
Return to the measured value display with <MODE>.
The stored data is retained.
1 Open the memory menu with <RCL>.
2 If necessary, scroll with <RCL> until CAL diSP is displayed.
3 Press <OK> to display the calibration data.
The following data is displayed consecutively for 10 seconds
each:
date, zero point, slope.
L
A
C
9
30
0
Day.Month
AutoCal