EasyManua.ls Logo

SCHOTT Lab 850 - Settings

SCHOTT Lab 850
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Lab 850
42
ba75890e01 11/2009
4.7 Settings
You can adapt the meter to your individual requirements. The settings
are done in the following menus:
z System settings (<OK
_>)
Baud rate (Baud)
Switch-off interval (t.Off)
Date (Day.Month)
Date (Year)
–Time (Time)
z Settings for calibration and measurement (<MODE
_>)
Number of the buffer set for pH calibration (Set 1 ... 16)
Display of the buffer during calibration
(pH nominal value or measured voltage value in mV)
Unit of the value for the slope (mV/pH or %)
Unit of the value for the zero point (mV, pH)
Resolution (Hi, Lo)
Temperature unit (°C / °F)
Calibration interval (Int.C [0 ... 999])
z Setting for automatic storage (<STO
_>)
AutoStore interval (Int.1)
z Setting for data download (<SC
_>)
Data download interval (Int.2)
Note
You can exit the setting menu at any time by pressing <MODE>.
Settings already modified and confirmed with <OK> are stored.

Related product manuals