SCHOTT Lab 850 - Measuring Ranges, Resolution, Accuracy

SCHOTT Lab 850
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lab 850 Technical data
61
ba75890e01 11/2009
Stand October 30, 2009Translation of the legally binding German version
7.2 Measuring ranges, resolution, accuracy
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the electrodes and buffer solutions has to be taken into
account additionally.
Measuring ranges,
resolution
Variable Measuring range Resolution
pH - 2.000 ... + 19.999
- 2.00 ... + 19.99
0.001
0.01
U [mV] - 999.9 ... + 999.9
- 2000 ... + 2000
0.1
1
T [°C] - 5.0 ... + 120.0 0.1
T [°F] + 23.0 ... + 248.0 0.1
Manual
temperature input
Variable Range Increment
T
manual
[°C] - 25 ... + 125 1
T
manual
[°F] - 13 ... + 257 1
Accuracy (± 1 digit) Variable Accuracy Temperature of the test
sample
pH / range *
- 2.000 ... + 19.999 ± 0.005 + 15 °C ... + 35 °C
- 2.00 ... + 19.99 ± 0.01 + 15 °C ... + 35 °C
U [mV] / range
- 999.9 ... + 999.9 ± 0.3 + 15 °C ... + 35 °C
- 2000 ... + 2000 ± 1 + 15 °C ... + 35 °C
T [°C] / temperature sensor
NTC 30 ± 0.1
PT 1000 ± 0.3
* when measuring in a range of ± 2 pH around a calibration point