6.0 Frequently Asked Questions
6.1 No Movement
Check Scientifica Control rack is properly connected according to the Control rack
User Manual and that the rack is turned on.
Ensure all cables and devices are correctly and securely connected as per the
relevant Control Rack User Manual and the rack is turned on.
Ensure that the correct device is selected using the device selection toggle switch on
the side of the Control Cube.
Ensure that the speed toggle switch is set to fast mode (and that a suitable speed
has been set in the LinLab software) and that the approach toggle switch is set to the
O position
Check all cables and connectors for damage or loose connections
6.2 Drift
Allow the system to warm up
With most electrical components on the rig needing a warm-up period, it is good
practice to allow a certain warming-up time after switching on the power of the set-up
(up to one hour).
Check all bolts are tightened including the following
The screws/bolts fixing the rotary base of the PatchStar down onto mounting
platform. At least 3 of the fixing holes on the rotatable base should be used. If using a
Post and Platform, then make sure the mounting platform is securely fastened with
the locking screw.
The 2 screws used for clamping the silver dovetail clamps on each stage are fixed
properly. Make sure the dovetails overlap on all stages before tightening the screws.
The 2 screws fixing the sliding bracket onto the vertical rotation plate should be firmly
tightened.
If the shallow or steep bracket is used, ensure that all 4 mounting screws are used.