❖ Two microprocessors for controlling instrument and processing test calculations
(Constant temperature, temperature control components, and two-dimensional bar
code scanning acquisition component).
❖ A thermal printer for printing results (printer components).
❖ 7.0-inch color capacitor multi-touch screen (display screen).
❖ Multiple selection features related to testing and results processing (operating
software).
2.5 Analyzer Function
❖ 7.0-inch touch screen, Android operation system with multi-language support.
❖ Single channel test, no cross contamination.
❖ Advanced optical inspection system, built-in 8 wavelength filters: 340, 405, 450, 505,
546, 600, 630, 850 nm
❖ Test methods: End point, rate, fixed time.
❖ Sample information storage can be satisfied with customer information storage
demand.
❖ Intelligent real-time quality control guaranteed accurate test results.
❖ Support for external mouse and keyboard (USB).
❖ Built-in thermal printer.
2.6 Scope of Application
The element RC is suitable for biochemical analysis of Li-Hep whole blood, Li-Hep plasma
and serum when used in conjunction with the available test rotors. It is for veterinary
medicine use only.
2.7 Executive Standard
❖ The main performance indicators of this product are designed and manufactured in
strict accordance with YY/T 0655–2008 "Dry Chemical Analyzer".
❖ The electrical safety of the product complies with the provisions of GB 4793.1–2007
(Safety requirements for electrical equipment for measurement, control and laboratory -
Part 1: General Requirement). The environmental test of the product meets the
requirements of GB/T14710–2009 (Environmental Requirements and Test Methods for
Medical Electrical Equipment).
❖ This product meets the special requirements of YY0648–2008 medical equipment
Section 2-101 of (Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use).
NOTE: Repeating any of the tests in GB 4793.1–2007 on the equipment may damage
equipment and increase the risk of danger!