5.4 Test Procedure Precautions
Analyzer
Use grounded power supply and supplied power cord.
Confirm the ambient temperature of the analyzer is 10ºC30ºC.
Do not disconnect power during operation.
Close the drawer when the analyzer is not in use.
Pinch hazard keep fingers clear when analyzer drawer is opening and closing.
Do not disassemble the analyzer.
Test Rotor
Do not use expired test rotors. Check expiration date on sealed pouch.
Store test rotors at 2ºC8ºC. Use gloves when handling test rotors, handle test rotors
by the edges.
After pipetting patient sample, place horizontally to prevent overflow.
Test rotors are single use; do not re-use.
Test rotors must be used within 10 minutes of removal from sealed pouch.
Analyze test rotor immediately after pipetting patient sample.
Sample
Whole blood should be used within 30 minutes after sampling to prevent hemolysis.
Do not place sample in the refrigerator or shake vigorously.