5.4 Test Procedure Precautions
Analyzer
❖ Use grounded power supply and supplied power cord.
❖ Confirm the ambient temperature of the analyzer is 10ºC–30ºC.
❖ Do not disconnect power during operation.
❖ Close the drawer when the analyzer is not in use.
❖ Pinch hazard – keep fingers clear when analyzer drawer is opening and closing.
❖ Do not disassemble the analyzer.
Test Rotor
❖
Do not use expired test rotors. Check expiration date on sealed pouch.
❖ Store test rotors at 2ºC–8ºC. Use gloves when handling test rotors, handle test rotors
by the edges.
❖ After pipetting patient sample, place horizontally to prevent overflow.
❖ Test rotors are single use; do not re-use.
❖ Test rotors must be used within 10 minutes of removal from sealed pouch.
❖ Analyze test rotor immediately after pipetting patient sample.
Sample
❖
Whole blood should be used within 30 minutes after sampling to prevent hemolysis.
❖ Do not place sample in the refrigerator or shake vigorously.