SECTION 3: INSTRUMENT INSTALLATION
3.1 Operating Environment
To ensure optimal performance, the instrument should be installed in accordance with the
following environmental conditions:
Elevation not to exceed 6500 feet (2000 m)
Temperature: 50ºF86ºF (10ºC30ºC)
Relative humidity: 30%70%
Barometric pressure: 860 hPa ~1060 hPa.
The analyzer should be installed according the following conditions:
Level, stable surface
Temperature and humidity control
Avoid direct sunlight or excessive heat
Avoid sources of vibration or electrical noise and interference (bench top centrifuge)
Do not install the instrument in following conditions:
Excessively high humidity, in the presence of corrosive gases, in dusty or unfiltered air,
or near strong electromagnetic
interference
Without circulating air or adequate ventilation
In direct sunlight or near other heat sources
Unsteady or tilted work surface
Near a centrifuge or other source of vibration or electrical noise
Requirements for power supply:
AC 240 V, 50 Hz, rated power: 120 VA
The instrument cannot use the same power supply socket with some high-power
equipment including centrifugal, refrigerator, oven, etc.
The power supply should earth well and connect the power supply with three-cores
wire; the voltage between the neutral line and earth line < 5 V.