SECTION 3: INSTRUMENT INSTALLATION
3.1 Operating Environment
To ensure optimal performance, the instrument should be installed in accordance with the
following environmental conditions:
❖ Elevation not to exceed 6500 feet (2000 m)
❖ Temperature: 50ºF–86ºF (10ºC–30ºC)
❖ Relative humidity: 30%–70%
❖ Barometric pressure: 860 hPa ~1060 hPa.
The analyzer should be installed according the following conditions:
❖ Level, stable surface
❖ Temperature and humidity control
❖ Avoid direct sunlight or excessive heat
❖ Avoid sources of vibration or electrical noise and interference (bench top centrifuge)
Do not install the instrument in following conditions:
❖
Excessively high humidity, in the presence of corrosive gases, in dusty or unfiltered air,
or near strong electromagnetic
interference
❖ Without circulating air or adequate ventilation
❖ In direct sunlight or near other heat sources
❖ Unsteady or tilted work surface
❖ Near a centrifuge or other source of vibration or electrical noise
Requirements for power supply:
❖ AC 240 V, 50 Hz, rated power: 120 VA
❖ The instrument cannot use the same power supply socket with some high-power
equipment including centrifugal, refrigerator, oven, etc.
❖ The power supply should earth well and connect the power supply with three-cores
wire; the voltage between the neutral line and earth line < 5 V.