__________________________________________________________________________
Vers. 20210510ENG Page 1
Product Manual
Veterinary Chemistry Analyzer
element RC

Questions and Answers

  • M
    michelle64Aug 15, 2025
    What to do if scil element RC Measuring Instruments shows mixing error?
    • T
      Tamara EstesAug 15, 2025
      If the scil Measuring Instruments displays a mixing error, verify that the sample is properly mixed. Alternatively, rerun the test using serum or plasma.
  • D
    dreyesAug 18, 2025
    What to do if scil element RC shows error code W3003?
    • T
      Terry ClarkAug 18, 2025
      If your scil Measuring Instruments shows error code W3003, this could be due to a sample distribution error caused by severe lipemia or a high HCT. To resolve this, centrifuge the sample and rerun the rotor using plasma or serum.
  • K
    Kathy CamposAug 22, 2025
    What to do if scil Measuring Instruments shows error code M5041?
    • A
      Andrew RichardsonAug 22, 2025
      If the scil Measuring Instruments displays error code M5041, it means the rotor is expired. Check the date setting of the analyser and rerun the test with a new rotor.
  • T
    Tracy WellsAug 26, 2025
    What to do if scil Measuring Instruments shows error codes E1012, E1024, W2061, or W2062?
    • C
      Connie ColemanAug 26, 2025
      If the scil Measuring Instruments displays error codes E1012, E1024, W2061, or W2062, it indicates that the rotor or QR of the rotor is damaged. Restart the analyser and rerun the test.
  • L
    Linda RochaAug 28, 2025
    What to do if scil Measuring Instruments shows error code W3001?
    • M
      Maria AllenAug 28, 2025
      If the scil Measuring Instruments displays error code W3001, it indicates a used rotor or damaged QC code. Rerun the test with a new rotor.
  • L
    Lauren LynchAug 31, 2025
    What to do if scil element RC shows error codes E1002, E1003, or E1023?
    • A
      Autumn SmithAug 31, 2025
      If the scil Measuring Instruments displays error codes E1002, E1003, or E1023, restart the analyser and rerun the test.
  • A
    April LaraSep 4, 2025
    What to do if scil element RC Measuring Instruments shows error code M5021 or M5022?
    • M
      michaelfowlerSep 4, 2025
      If the scil Measuring Instruments displays error codes M5021 or M5022, check the printer connection and ensure the printer has paper loaded correctly.
  • M
    Mrs. Kayla CampbellSep 7, 2025
    What to do if scil Measuring Instruments shows error code M5002?
    • T
      Tara FarmerSep 7, 2025
      If the scil Measuring Instruments displays the error code M5002, check the printer to ensure paper is properly loaded.
  • W
    wjohnsonSep 12, 2025
    What to do if scil element RC shows error code M4002?
    • J
      James ScottSep 12, 2025
      If the scil Measuring Instruments displays error code M4002, it means that the current rotor is not supported due to a low software version. Check the date setting of the analyser. Please contact the technical service and update the device.