5.2.2 Pipette patient sample
1. Use the supplied 100μl pipette and
disposable tip to pipette patient
sample into the sample port.
Fill the sample well to the fill line.
2. Pipette tips are single use, do not re-
use to avoid sample contamination.
3. Hold the test rotor horizontally by the
edges when insert into the analyzer
drawer.
CAUTION
Do not mark or damage QR code on test rotor.
Do not touch pipette tip.
Place pipette tip approx. 1/16 in (23 mm) below the surface of the sample when
pipetting.
Keep the pipette plunger pressed down until the pipette tip is removed from the sample
port on test rotor.
Clean any sample on test rotor or around sample port before placing test rotor in
analyzer.
Fill line
Fill port