EasyManuals Logo

scil element RC User Manual

scil element RC
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
__________________________________________________________________________
Vers. 20210510ENG Page 22
5.2.2 Pipette patient sample
1. Use the supplied 100μl pipette and
disposable tip to pipette patient
sample into the sample port.
Fill the sample well to the fill line.
2. Pipette tips are single use, do not re-
use to avoid sample contamination.
3. Hold the test rotor horizontally by the
edges when insert into the analyzer
drawer.
CAUTION
Do not mark or damage QR code on test rotor.
Do not touch pipette tip.
Place pipette tip approx. 1/16 in (23 mm) below the surface of the sample when
pipetting.
Keep the pipette plunger pressed down until the pipette tip is removed from the sample
port on test rotor.
Clean any sample on test rotor or around sample port before placing test rotor in
analyzer.
Fill line
Fill port

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the scil element RC and is the answer not in the manual?

scil element RC Specifications

General IconGeneral
Brandscil
Modelelement RC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals