SD GlucoNavii® NFC BLOOD GLUCOSE MONITORING SYSTEM
Maintenance and Troubleshooting
59
[Lo message]
t Blood glucose may be lower than
the measuring range of the system.
See chapter 3 “4. Understanding
Test Results”.
[Internal Error Message for a
meter]
t Turn o a meter. Then turn on
the meter again. If there is still
error message, please contact SD
Biosensor, Inc. TEL: +82-31-300-0400.
[Strip Error]
t Defective test strip or the test strip
is damaged or inserted improperly.
Discard this test strip and test again
using new test strip. See Chapter 3 “2.
Performing a Blood Glucose Test”.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SD Biosensor SD GlucoNavii NFC and is the answer not in the manual?

SD Biosensor SD GlucoNavii NFC Specifications

General IconGeneral
BrandSD Biosensor
ModelSD GlucoNavii NFC
CategoryBlood Glucose Meter
LanguageEnglish