80 •
Activating/deactivating key lock Key lock enables you to prevent undesired settings being made to the
multifunctional display once the device has been switched on.
NOTE:
No separate symbol for activated key lock appears in the
multifunctional display.
Key lock cannot be activated while you are navigating in the menu.
1. Ensure that the device is switched off.
2. Press the Start key of the multifunctional display.
3. Wait until the display
seca appears.
4. Press the menu key of the multifunctional display.
Key lock is switched on.
5. To cancel key lock again, restart the device by pressing the key of
the multifunctional display.
6. Wait until the display
seca appears.
7. Press the menu key of the multifunctional display.
Key lock is switched off.
Weighing babies/toddlers (2-in-1) The 2 in 1 function can be used to determine the weight of babies and tod-
dlers. The child is held in an adult's arms during weighing to do so. Proceed
as outlined below.
1. Ensure that there is no load on the scale.
2. Switch on the device (see "Switching on the device" on page 79).
3. Ask the adult to stand on the scale.
The adult's weight is displayed.
4. Press the
2 in 1 key.
The weight is saved.
0.00, the symbol (non-verifiable function) and the message "NET"
appear in the display.
ATTENTION!
Incorrect measurement as a result of initial weight changing
If weighing of the child commences with a different initial weight, it will
not be possible to determine the child's weight correctly.
Ensure that the child is always weighed with the adult with whom the
initial weight was determined.
Ensure that the adult's weight does not change - as a result of their
taking off a piece of clothing, for example.
5. Ask the adult to get onto the scale holding the child.
The child's weight is displayed.
The symbol, the symbol and the messages "HOLD" and "NET" are
displayed.
6. Ask the adult to get off the scale with the child.
7. To deactivate the
2 in 1 function, press the 2 in 1 key or switch off the
scale.