10.MAINTENANCE
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor servicing
►
Have servicing and repairs carried out exclusively by seca service or
by an authorized service partner.
► You can find service partners in your area or by
sending an e-mail
11.TECHNICAL DATA
11.1 General technical data
... the display "Er :6 :82" appears?
An error occurred during the measuring operation.
- Ask the patient to keep still and repeat the measuring operation
• If the error message recurs:
-inform seca Service
... the display "Er :6 :83" appears?
An error occurred during the reference measurement.
- Step off the platform
• Interference caused by reflection
- Remove objects from the immediate vicinity of the device (see "Setting up
the device" on page 74).
- Ask people to keep a reasonable distance from the device (see "Setting
up the device" on page 74).
• Interference cause by other ultrasonic emitters
- Increase the distance from other ultrasonic emitters (see "Setting up the
device" on page 74)
... the display "Er :6 :84" appears?
• The ambient temperature of the device is too high or too low.
- Set up the device in an ambient temperature between +10 °C and
+40 °C
- Wait for around 15 minutes until the device has adapted to the ambient
temperature
• The temperature sensor is faulty.
- Inform seca Service
Problem Cause/solution
General technical data
Dimensions
•Depth
•Width
•Height
466 mm
434 mm
2270 mm
Net weight 16.5 kg
Ambient conditions, operation
• Temperature
•Air pressure
• Humidity
+10° C to +40° C (50 °F to 104 °F)
700 hPa - 1060 hPa
30 % to 80 %, no condensation
Ambient conditions, storage
• Temperature
•Air pressure
• Humidity
-10° C to +65° C (14 °F to 149 °F)
700 hPa - 1060 hPa
0 % to 95 %, no condensation