10.MAINTENANCE
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor servicing
Have servicing and repairs carried out exclusively by seca service or
by an authorized service partner.
You can find service partners in your area or by
sending an e-mail
11.TECHNICAL DATA
11.1 General technical data
... the display "Er :6 :82" appears?
An error occurred during the measuring operation.
- Ask the patient to keep still and repeat the measuring operation
If the error message recurs:
-inform seca Service
... the display "Er :6 :83" appears?
An error occurred during the reference measurement.
- Step off the platform
Interference caused by reflection
- Remove objects from the immediate vicinity of the device (see "Setting up
the device" on page 74).
- Ask people to keep a reasonable distance from the device (see "Setting
up the device" on page 74).
Interference cause by other ultrasonic emitters
- Increase the distance from other ultrasonic emitters (see "Setting up the
device" on page 74)
... the display "Er :6 :84" appears?
The ambient temperature of the device is too high or too low.
- Set up the device in an ambient temperature between +10 °C and
+40 °C
- Wait for around 15 minutes until the device has adapted to the ambient
temperature
The temperature sensor is faulty.
- Inform seca Service
Problem Cause/solution
General technical data
Dimensions
•Depth
•Width
•Height
466 mm
434 mm
2270 mm
Net weight 16.5 kg
Ambient conditions, operation
Temperature
•Air pressure
Humidity
+10° C to +40° C (50 °F to 104 °F)
700 hPa - 1060 hPa
30 % to 80 %, no condensation
Ambient conditions, storage
Temperature
•Air pressure
Humidity
-10° C to +65° C (14 °F to 149 °F)
700 hPa - 1060 hPa
0 % to 95 %, no condensation