100 •
... no beeps can be heard?
Beeps are not activated
- Activate function (see "Activating/deactivating beeps (bEEP)" on page 91)
Ultrasound measurement is deactivated
- Restart the device
Volume set to zero
- Increase volume
Loudspeaker is faulty
- Inform seca Service
... the measured results are not
announced?
Announcement of measured results is not activated
- Activate function (see "Activating/deactivating announcement of
measured results (reSUL)" on page 86)
Ultrasound measurement is deactivated
- Restart the device
Volume set to zero
- Increase volume
Loudspeaker is faulty
- Inform seca Service
... the request to take the printed
ticket is not announced?
The function is not activated
- Activate function (see "Activating/deactivating the announcement “Take
your printed ticket” (Print)" on page 86)
Ultrasound measurement is deactivated
- Restart the device
Volume set to zero
- Increase volume
Loudspeaker is faulty
- Inform seca Service
... one segment of the multifunctional
display is on either continuously or
not at all?
The corresponding point has a fault.
- Inform seca Service
… the display "StOP" appears?
During weighing:
Maximum load has been exceeded.
- Remove the load from the scale
When setting up a wireless group:
Wireless channel setup is complete.
- Switch on devices to be integrated in the wireless group (see "Setting up
a wireless group (Lrn)" on page 95)
The display "tEMP" appears.
The ambient temperature of the scale is too high or too low.
- Set up the scale in an ambient temperature between +10 °C and +40 °C
- Wait for around 15 minutes until the scale has adapted to the ambient
temperature.
... the multifunctional display is no
longer reacting to keys being
pressed?
Key lock is active.
- Switch off key lock (see "Activating/deactivating key lock" on page 80)
Device is in an undefined state following implausible input.
- Take power supply unit out of the socket
- Wait approx. 1 minute
- Plug the power supply unit into the socket; the scale and the
multifunctional display will switch on automatically