10.SERVICING
The product must be set up carefully and serviced
regularly. Depending on how frequently the product is
used, we recommended servicing at intervals of 3 to 5
years.
NOTICE!
Incorrect measurements as a result of poor
servicing
Have servicing and repairs carried out
exclusively by seca Service or by an
authorized service partner.
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
The display "Er:[No.]:11"
appears.
The scale has too high a load or too high a load in
one corner.
-Take the load off the scale or distribute the
weight more evenly
-Restart the scale
The display "Er:[No.]:12"
appears.
The scale has been switched on with too high a
load.
-Remove the load from the scale
-Restart the scale
The display "Er:[No.]:16"
appears.
The scale was caused to oscillate, the zero point
could not be determined.
-Restart the scale
The Enter key
(send/print) is pressed
and the display
"Er:[No.]:71" appears.
No data transmission possible, wireless module
deactivated.
-Activate wireless module (see "Activating the
wireless module (SYS)" on page 103)
The Enter key
(send/print) is pressed
and the display
"Er:[No.]:72" appears.
No data transmission possible, no wireless group
set up.
-Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 103)
Problem Cause / Solution