Safety information • 63
English
WARNING!
Risk of slipping
Ensure that the weighing platform is dry
before the patient steps onto it.
Ensure that the patients feet are dry before he
or she steps onto the weighing platform.
Make sure that the patient steps onto the
weighing platform slowly and securely.
Preventing device
damage
ATTENTION!
Damage to device
Ensure that no liquids enter the device. They
can damage the electronics.
Switch off the device before disconnecting the
power supply unit from the mains socket.
Disconnect the power supply unit from the
mains socket if you intend to not use the
device for a longer period of time. Only this
way it can be ensured that the device is cur-
rentless.
Make sure not to drop the device.
Do not expose the device to any impacts or
vibrations.
Perform function controls regularly as
described in the relevant section in this
document. Do not operate the device if it is
damaged or not working properly.
Ensure that there is no heat source in the
immediate vicinity. Do not expose to direct
sunlight. The excessive temperature could
damage the electronics.
Avoid rapid temperature fluctuations. When
the device is transported so that a tem-
perature difference of more than 20 °C
occurs, it must stay turned off for at least 2
hours before it can be turned on again. Other-
wise, condensation water will form which can
damage the electronics.
Caustic detergents may damage the surfaces.
Only use a soft cloth dampened with mild
soapsuds to clean the surfaces of the device.
Only use disinfectants suitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.