Service manual Description of faults
21.11.08- Winterberg/Jen
1(2)
30-34-00-588 Index: d
Please refer to the operating instructions to make sure that scale malfunctions are not
caused by operation faults. Should any faults occur, the following table will help you to
identify the cause of the fault.
Fault description Possible cause Remedy
Scale does not start Battery not inserted
or flat
Check the batteries
Voltage supply
defective
Check the voltage at the relevant electronics connections, correct
the interruption
Bus connecting
cable defective
Check the plug-in connections, check the cable for interruptions,
replace if necessary
Keyboard defective
or not connected
Check the cable connections, check cable and keyboard for
interruptions, replace if necessary. (If possible please state the
number of the keyboard when ordering a replacement.)
Bus fault Disconnect power supply unit and battery from the scale, wait for 60
seconds, then reconnect.
Electronics
damaged by a
wrong type of power
supply unit
Replace the electronics. (Warning: The damaged electronics can
draw so much current that the power supply unit is also damaged.)
Scale repeatedly shows
no weight or an incorrect
weight
Packaging /
transport locking
device not removed
completey
Check the scale
Scale not correctly
adjusted
Recalibrate the scale
Sensor damaged Replace the sensor and readjust the scale
Lever system
damaged or load
transmission to
sensor interrupted
Check the base frame (and tie rod if present) for damage or incorrect
assembly
Scale remains in the
segment test or
measured values
fluctuate heavily
Short-circuit
between sensor
cable and housing
Look where the insulation is damaged (using a multimeter if
necessary) and repair it.
Display incomplete Display defective Replace the electronics
Scale does not respond
to keystrokes
Keyboard defective
or not connected
Check the cable connections, check cable and keyboard for
interruptions, replace if necessary. (If possible please state the
number of the keyboard when ordering a replacement.)
Faults occur during the
weighing operation
Display:
Er:x:no
Electronics detecting a fault.
X : Module number
No Fault number
Er:X:10
Sensor defective or
connected
incorrectly
Check the sensor and replace it if necessary (readjustment required
after replacement).
Er:X:11
Sensor defective,
supply lines
damaged or
electronics defective
Check supply lines, check sensor and replace as required
(readjustment required after replacement)
If several load cells are fitted, check the allocation (colour codes)
Replace the electronics
Er:X:12
Initial load too high Check the initial load
Check the load levers (jammed or bent), sensor damaged, replace
and readjust if necessary