Service manual Description of faults
21.11.08- Winterberg/Jen
2(2)
30-34-00-588 Index: d
Fault description Possible cause Remedy
Er:X:14
Fault in the kg value
calculation
Wrong coefficients in the module’s ROM, reprogram and readjust the
scale
Er:X:15
Recalibration fault Reference weight outside the permitted range or operating fault
Er:X:21
Fault in battery
voltage
measurement
Battery voltage much too low < 2.5 V or electronics defective
-> Replace the electronics
Er:X:22
Temperature
measurement error
Replace the electronics
Er:X:30
Data transmission
interrupted.
Check the bus connection, remove external sources of faults.
Er:X:32
Command buffer
overflow fault
Check the bus cabling, eliminate the EMC fault, de-energize the
electronics for a short instant (60 secs)
Er:X:33
To many modules
connected to bus
Reduce the number of modules connected to the bus.
Er:X:40
No data on
EEPROM / data
deleted
Readjust the scale or reload the scale adjustment data to the
EEPROM via the service interface
Er:X:41
Incorrect EEPROM
access
One of the modules connected accesses the memory in an incorrect
way. Determine the faulty module and replace it.
Er:X:50
Software checksum
not correct
Fault immediately after switch-on -> wrong checksum in the ROM
-> Replace the electronics
Er:X:51
Menu fault Fault when operating the menu functions (impermissible values or
similar)