Maintenance seca
05.10.12/RLEU/ARI/MRE 13/38 30-34-00-812b
been pressed.
Press the tare/hold key for more than 1.5
seconds.
"dec" appears on the display (flashing)
and then the current weight (0.0 - also
flashing).
Place the test weight on the scale. The measured weight of the test weight
appears on the display (flashing).
Decide whether to decrement or to
increment:
Incrementing:
Press the tare/hold key for more than 1.5
seconds.
Decrementing: No action required as the
scale already is in dec mode.
Incrementing mode is activated; "inc"
appears on the display.
No output on the display.
Decrementing or incrementing:
gain briefly press the tare/hold key (for
less than 1.5 seconds).
The measured weight is decremented or
incremented by 10 g. Please remember
that pressing the key does not always
change the value on the display due to
the low display resolution. Only the
rounded value is output.
Repeat this step until the measured
weight matches the test weight (format:
xxx.0 kg).
The display indicates the test weight
(format: xxx.0 kg).
To correct this rounding error, decrement
or increment the value five times by 10 g.
The display indicates the exact test
weight (format: xxx.0 kg).
Press the tare/hold key until the display
shows “Sto”.
The display reads “Sto” and the scale
switches off automatically.