METIS M311 / M322 / H311 / H322 (17-pin)
SensorTools Software
30
7.2.2 Control Window
Measuring mode 2C means that the
2-color temperature is displayed.
In the measured temperature selec-
tion field it can be changed to
measuring temperature of chan-
nel 1 or channel 2.
The setpoint of the PID controller
is always displayed.
If a ramp time is defined, the set-
point can always display the updat-
ed value when the buffer mode with
status information is selected in the
data collection settings (see
7.2.3.4).
The internal device temperature
(DT) depends on the ambient temperature and the device self-heating.
- M3 models are permissible up to maximum 80°C.
- H3 models are permissible up to maximum 60°C.
- The laser targeting light is deactivated at a device temperature from 60°C, the camera module of
M3 devices from 55°C to prevent its overheating). Below that temperature it can be used again.
The status of the digital inputs / outputs is indicated by an LED symbol:
- Input (IN) or output (OUT) active
- Input (IN) or output (OUT) inactive
- Input (IN) or output (OUT) not activated (no function)
Adapt parameters: Parameter changes are immediately transferred to the pyrometer.
- Device parameters
- The response time t
90
(see 6.6) can be adapted to the measurement conditions.
Selecting "Minimum" the pyrometer works with its shortest possible response time of <1 ms (M3)
or <80 µs (H3). Under "User definded" times between 1 ms (M3) or 0.1 ms (H3) and 10 s can be
entered.
- Storage mode (see 6.7): switches on (and off) the peak picker and allows clear settings for the
activated peak picker.
- Measurement parameters: The displayed measuring parameter depends on the measuring
mode. In case of ratio temperature measurements, the emissivity slope can be set (see 6.1); in the
case of 1 channel measurements, the emissivity input field is available first.
- The setting field can be expanded to display the measurement parameters for all channels.
- To display further sub-parameters, each individual parameter can be extended
- Emissivity slope (see 6.1): Adjustment of the surface
properties in 2-color mode.
- Switch-off level (see 6.5): Input when the measurement
is turned off by low signal strength.
- Switch-off verification (see 6.5.1): Input when the
switch-off is actually made.
- Emissivity (see 6.2): Adjustment of the surface proper-
ties in 1-color mode.
- Transmittance (see 6.3): Transmittance of a window
- Spot size fill factor (see 6.4): For 1-color measure-
ments, for targets smaller than the spot size.
Quick adjustment of parameters
: Changes the currently selected
parameter (e.g. click with the mouse in a field or select it us-
ing the Tab key in sequence of the fields).
Controller parame-
ters (optional)
Laser targeting light on-
off / camera settings
Status of the digital
inputs / outputs
Measurement pa-
rameters quick save
Parameter quick
adjustment
Set point temp.
(devices with PID
controller)
Measurement parame-
ters (extendable)
Motor focus adjustment
(optional)
Show measuring
parameters of all
channels
Further settings
for measuring
parameters