EasyManua.ls Logo

SeQual INTEGRA E-Z - Starting the Integra E-Z; Start-Up Sequence; Start-up Sequence for OM Units; Start-up Sequence for Non-OM Units

SeQual INTEGRA E-Z
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual
Integra E-Z Oxygen Concentrators
Chart SeQual Technologies Inc. Page 9 of 34 P/N 2917
Flow Rate
Shown?
Green
Light
Yellow
Light
Red
Light
Audible
Alarm
Unit
State
Operating Condition
No Off Off Off Off
Not
Energized
Unit not running / Flow
Indicator Display Not Lit
No Off Off Off Off Energized
Unit not running / Flow
Indicator Display Lit
2.5. Starting the Integra E-Z
2.5.1. Push the On/Off button and verify the unit Start-Up Sequence
Allow the unit to warm-up and stabilize (approximately 2 minutes but allow up to 10 minutes).
Adjust the flow using "+" and "" buttons. A shorter warm-up will occur at lower flow rates.
NOTE:
The Integra E-Z flow indicator sets the target flow rate desired by the patient. The
installation of humidifiers or other devices can cause a back pressure on the unit
therefore it is recommend that the flow rate be set prior to attaching accessories to
the concentrator. After an initial startup or after a power interruption the unit may
automatically reduce the flow rate briefly before returning to the set flow rate.
2.5.2. Start-up Sequence OM units
On initial start-up Integra E-Z units light the Green, Yellow and Red LED’s with a 3 second
beep followed by an intermittent beep for the remainder of the warm-up. As concentration
increases to ~70%, the Red LED turns off and the Yellow and Green remain on. When the
concentration reaches normal operating range (above 85%), the Yellow LED will turn off as
well as the intermittent beep. The Green LED will stay lit indicating normal operation.
2.5.3. Start-up Sequence Non-OM units
On Non-OM start-up Integra E-Z units will light the Green, Yellow and Red LED’s with a 3
second beep indicating all alarms indicators are functioning followed by the green LED turning
on. The table below shows operating conditions for Integra E-Z Oxygen Concentrators:
Flow Rate
Shown?
Green
Light
Yellow
Light
Red
Light
Audible
Alarm
Unit
State
Operating Condition
Yes On On On
On for 3
seconds
Started
Running
Unit is Starting
Yes On On On
Periodic
Beep
Running
Unit is warming up*
Yes On On Off
Periodic
Beep
Running
Yes On Off Off Off Running
Normal Operating
Condition
Yes Off On Off
Periodic
Beep
Not
Running
Unit turned
On/Off/On Quickly
* These conditions are displayed only on units equipped with oxygen monitor.

Related product manuals