28
Display 4 - 1,000 ppm adjustmentDisplay 4 - 1,000 ppm adjustment
Display 4 - 1,000 ppm adjustmentDisplay 4 - 1,000 ppm adjustment
Display 4 - 1,000 ppm adjustment
Now release the
1,000 ppm methane CH1,000 ppm methane CH
1,000 ppm methane CH1,000 ppm methane CH
1,000 ppm methane CH
44
44
4
test gas test gas
test gas test gas
test gas from the
SPE 3SPE 3
SPE 3SPE 3
SPE 3
test set.
wait until the display has stabilised
confirm the adjustment with the
on/off button (
OKOK
OKOK
OK appears in the
LCD display)
turn off the test-gas feed
press the cursor-up button to
move to the next display
NB: change the test set !NB: change the test set !
NB: change the test set !NB: change the test set !
NB: change the test set !
Display 5 - 1.00 vol.% adjustmentDisplay 5 - 1.00 vol.% adjustment
Display 5 - 1.00 vol.% adjustmentDisplay 5 - 1.00 vol.% adjustment
Display 5 - 1.00 vol.% adjustment
Now release the
1.00 vol.% methane CH1.00 vol.% methane CH
1.00 vol.% methane CH1.00 vol.% methane CH
1.00 vol.% methane CH
44
44
4
test gas test gas
test gas test gas
test gas from the
SPE 1SPE 1
SPE 1SPE 1
SPE 1
test set.
wait until the display has stabilised
confirm the adjustment with the
on/off button (
OKOK
OKOK
OK appears in the
LCD display)
turn off the test-gas feed
press the cursor-up button to
move to the next display