5-3-4. Measuring the DEAD Weight of Weighing Scale
5-3-5. Inputting Max Output (Rated Output Voltage / mV)
After mV/V is displayed, Check the Rated output value of Load cell.
(Refer to the load cell label, or Test Report). And Press key to save and move to
next step.
When DEAD displays, Press key, then indicator will calculate dead weight of
scale part automatically (While this process, there should be nothing on the scale part.)