Sewha SI 410 - CALIBRATION ERROR TROUBLESHOOTING; Error handling during calibration

Sewha SI 410
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 23 -
If input wrong value, there will display Err-01, please go back to Setting Capacity of
Load Cell.
After recheck the label of load cell and retry the process.
Calculated span value will be displayed. Then
press key to finish the calibration
step.
In case of multiple pieces of load cells are connected, the rated output will be
same as single Loadcells. (Because plural load cells are connected with parallel
connection, the sum of rated output voltage is same as single load cells rated output)
Due to some variation between “State output rate” and “Real Output rate” of load
cell, there might be some weight difference after finishing calibration.
If you want to make more precise weighing process, please measure real output rate of
load cell and input the measured value. Then the weight measurement will be more
precise than before.