Port 2 Modbus communication
data SWAP
00 : Basic
01 : Customizing
Ethernet Communication mode
00: Simplex / Stream Mode
01: Duplex / Command Mode
02: Duplex / Command Mode (Compatible with
SI4100)
04: Modbus(RTU)
Ethernet Format under Stream
Mode
00 : Format 1 (18byte)
01 : Format 2 (21byte)
02 : Format 3 (17byte)
03 : Format 4 (22byte)
04 : Format 5 (15byte)
Ethernet transference under
stream mode
00 : Countinuously
01 : Single time on every steady state
02 : Single time at the first steady point
03 : Single time output after weighing finish
04 : When input Print key
Ethernet “Check-Sum” under
command mode
Ethernet Modbus LCB/MSB
setting
Remark: If BCD-OUT option is needed, F250 must set as 00
00 : Forward
01 : Reverse
Analog Output Applying
Weight Range
00 : Absolute number(-&+)
01 : Positive number(only +)
00 : Within 10% of Max Capacity
01 : Within 20% of Max Capacity
02 : Within 50% of Max Capacity
03 : Within 100% of Max Capacity
BCD IN Enter method
(Part Number)
00 : Disuse
01 : Enter the Unit digit, and Tenth digit with
dividing
02 : Enter the Unit digit, and Tenth digit
without dividing
03 : Absolute value