EasyManua.ls Logo

Shining 3D Transcan-C - Scanner Calibration; Calibration Precautions, Operation, and White Balance

Shining 3D Transcan-C
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Shining 3D Transcan-C_User Manual_V1.4.2
18
4. Calibrate the Scanner
4.1. Precautions and Use
Calibration is the process to ensure the device will scan with the optimal accuracy and scan
quality.
This customer calibration is not required daily often. Only when the first time after scanner and
software are installed. Or, when you are in the following situations, you should calibrate the device:
When the scanner is used for the first time or after long time without using.
When there is strong vibration during the transportation.
When alignment mistake or failure frequent appear during the scanning.
When scanning data is incomplete and quality is much worse during the scanning.
4.2. Operation
After installation, when you open the software for the first time, choose device type and it will
enter the calibration interface automatically. If there is no calibration data, click “quit”, the
software will prompt “No calibration data, please calibrate”.
Different calibration boards are used for scanning ranges of 300mm and 150mm. Select the
corresponding calibration board as shown in the calibration interface.

Related product manuals