3.2 Replacing the water trap
Overview
1
Retaining ring, hose connection
2
Screws
3
Retaining ring, water trap
Prerequisites
Analyzer in “Stand-by
System flushed for 10 minutes
Electronic module housing removed
Procedure
On the water trap
1.
Remove the hose connection (DN 2/4) by pressing retaining ring 1 on the hose
connection down.
2.
On the bracket: Loosen screws 2.
3.
Press retaining ring 3 at the top of the water trap down.
4. Slide the water trap with bracket down and remove.
5. Remove the water trap from the bracket by pressing the bracket down onto the
connection.
6. Note the date of installation on the new water trap.
7. Insert the new water trap together with the bracket into the connection.
NOTICE Observe the flow direction of the water trap. Labeling “GAS-IN” must
point upwards.
8. Mount the water trap with “GAS-IN” facing upwards.
9. Push the bracket upwards and tighten the screws.
10. Reattach the hose connection.
Finishing work
1. Reattach the electronic module housing.
2. Reset stand-by and maintenance signal.
MAINTENANCE 3