17
9. OPERATION OF THE BED
• Ensure the bed and all accessories are at room temperature.
• Ensure the bed has been cleaned and disinfected (see section 13).
• Ensure the brakes on the castors at the head end of the bed have been applied.
• Using the handset ensure the bed is level (see section 9.6 for handset operation).
If the electrical functions do not operate, ensure the handset has been ‘unlocked’ (see
section 9.6 for handset operation).
9.1 Operational Limits
See section 16.2.
9.2 Installation / Preparing for Use
Prior to operating the bed for the first time the following simple checks must be
performed:
• Thebedistobeleftinitslowestpositionwhenthepatientis
unattendedinordertoreducetheriskofinjuryduetoafall.
• Beforeoperatingthebedensurethepatientispositioned
appropriately ensuring all limbs are clear of moving parts.
• Themainsplugisthedisconnectdeviceforthemeansofisolatingthe
bed from the mains supply, the plug must be accessible at all times.
• Ensurethemainscableispluggedintoanappropriatepowersourceat
all times.
• Ensuretheelectricalcablesarenotintension,payingparticular
attention to the mains cable when the bed travels up/down.
• Ensurethatallcablesandthehandsetareclearofallmovingpartsto
prevent damage to the electrical components.
• Inappropriatehandling/positioningofthemainscablecouldcause
kinkingorshearingofthecablewhichmayleadtoexposedlivewires-
Risk of electrocution.
• Precautionsaretobetakenwhenroutingcablesfromexternal
equipment around the bed to ensure that they do not become
squeezed, trapped or damaged.
• ACEmarkedextensioncablemustonlybeusedwhenitisnotpossible
toreachawallsocketwiththeequipmentmainscable–ContactSidhil
Ltd. for detail with regards to the safe use of extension cables.
Warning