34
16.2 Electrical Data
Voltagein: 230V±10%,~50/60Hz.
Current in: 1.25A
Mainscablefuse: 3-5A
Dutycycle:* 10%
2 mins of continuous use followed by 18 mins not in use.
*Electricallyoperatedbedsareintendedtobeoperatedintermittentlyratherthancontinuously.Ifthebedisoperated
continuously for up to 2 minutes it must then be left for at least 18 minutes before reuse to allow the electrical system to cool
sufficiently. If the bed is continuously used for an extended period of time and it exceeds the duty cycle the control
box may become temporarily disabled or irreparably damaged.
Safetystandards: IEC60601-1:2005
IEC60601-1-2:2007
IEC60601-2-52:2009
TheSoliteUKBediscomplianttotherequirementsof60601-1-2:2007,theRF
emissions are very low and are not likely to cause any interference to nearby electronic
equipment, however interference to sensitive equipment is possible.
Qualified users of the Solite UK Bed can help prevent electromagnetic interference
by maintaining a suitable distance between portable and mobile RF communications
equipment (transmitters). If the Solite UK Bed or any alternative equipment is found to
be operating abnormally mitigation measures are to be taken, such as the separation
distances being increased and / or the products being reorientated.
Applied parts: Mattress platform
Profiling sections
Upper half of head / foot ends
Handset
Liquidingressprotection: IPX4-Splashresistant
Battery backup: Individual batteries non replaceable
(SOLITE/UK/4/BB only)
Noise level: 59dB(A) max
Service life: 5 years
Environmental conditions:
Operational Limits Transportation / Storage Limits
Ambient Temperature +5ºCto+40°C -10°Cto+50ºC
Humidity 2 0 % - 9 0 % a t 3 0 ° C
non condensing
2 0 % - 9 0 % a t 3 0 ° C
non condensing
Atmospheric Pressure 800 to 1060hPa 800 to 1060hPa
Altitude ≤2000m ≤2000m