120 ADVIA 2400 Operator’s Guide
Checking other vertical pumps for leaks
Use this procedure to test for leaks from other vertical pumps: the sampling wash pump
(SCP), dilution wash pump (DCP), dilution discharge pump (DOP), reagent dispensing
pumps (RP1 and RP2), and reagent wash pumps (RWP1 and RWP2).
1. To determine if any of these pumps are leaking, inspect the upper portion (1)
(cylinders, L-ring holders, tubes and fittings) of the vertical pumps.
Figure 5-11. Typical vertical pump
2. If the pump is leaking, you must replace the pump seal. To replace the pump seal,
call your local technical support provider or distributor.
Checking the horizontal pumps for leaks
Horizontal pumps consist of the dilution cuvette wash pumps, DWP1, DWP2, DWP3 and
DWP4; the reaction cuvette wash pumps, WP1, WP2, and WP3; and reaction cuvette
detergent pumps DTP1 and DTP2).
1 DWP1
2 DWP2
3 DWP3
4 DWP4
5 WCV2
6 WCV1
7 WP1
8 WP2
9 WP3
10 DTP1
11 DTP2
Figure 5-12. Horizontal pumps