158 ADVIA 2400 Operator’s Guide
4. If not centered, call your local technical support provider or distributor.
5. At the Operation Panel, select
Initialize to return the probes back to home (over the
wash cups).
6. At the Operation Panel, select
Prime, PRIME 2, and then Execute to ensure proper
water flow through the probe.
NOTE
Make sure that no water is leaking from the joint connector (4).
Replenishing the RRV (reaction) bath oil bottle
Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1 ISE buffer bottle
2
RRV (Reaction) bath oil bottle
3
Isotonic saline diluent bottle
4 Cuvette detergent bottle
5 Cell conditioner bottle
Figure 5-33. RRV bath oil bottle
CAUTION
Do not attempt to clean the RRV bath oil bottle (1) with water; RRV bath oil and water
do not mix.
1. Unscrew the filter cap (
3) at the front of the RRV bath oil bottle (1), then pull up the
tube with the filter.
2. Disconnect the RRV bath oil bottle level sensor connector (
2), then turn the
connector counter-clockwise and pull it out.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
3. Remove the RRV bath oil bottle (
1).
CAUTION
Ensure that level-sensor connector (2) does not get wet, to avoid damaging it.