Maintenance 137
If you are cleaning other bottles, wait to perform this step for all fluid lines.
a. At the Operation Panel, select the
Prime button.
b. At the PRIME Settings dialog box, select
Prime 2 and then type 10 or more for
the number of times in all fields.
c. Select
Execute.
Cleaning the chiller filter
Materials required:
• Vacuum cleaner
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE
Access the chiller filter (located on the right inside bottom shelf of the analyzer cabinet)
through the panel door on the right side of the analyzer.
1. On the right side of the analyzer, push and release the panel door to gain access to the
chiller unit.
2. Locate the filter and slide it out of the analyzer.
3. Using a vacuum cleaner, remove the dust from the filter.
4. If the filter requires further cleaning, perform the following steps:
a. Wash the filter under running water.
b. Dry the filter before replacing it.
5.
Slide the filter back in place and close the panel on the right side of the cabinet.