134 ADVIA 2400 Operator’s Guide
Cuvette wash and cuvette conditioner usage
CTT and RTT
Solution Approximate
Volume Used
During Wash 2
Number of
Aspirations
Total number of
WUD cycles for
Wash2
10% Cuvette Wash
DPP = 3.6 mL
RPP1 = 29 mL
RPP2 = 29 mL
DPP = 0.36 mL
RPP1 = 2.9 mL
RPP2 = 2.9 mL
1220 (approx)
System solutions
Solution Dilution by
system
Volume
dispensed
per WUD
cycle
Volume of
undiluted
solution
used per
WUD cycle
Total volume
used for
Wash2
undiluted
Cuvette Wash 1:10 with water
600 μL 60 μL
44 ml
Cuvette Conditioner 1:40 with water
600 μL 15 μL
11 ml
NOTE
The total volume of cuvette wash and cuvette conditioner used by the system can vary
slightly from the volumes provided in the tables above. This is normal behavior.
Cleaning and replenishing the dilution bottle
Materials required:
• Deionized Water
• Physiological saline (0.9% NaCl)
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE
The dilution bottle may be cleaned when it is refilled, but must be cleaned at least once a
month.