146 ADVIA 2400 Operator’s Guide
Cleaning the pure-water bottle filter
Materials required:
• Ten 10R filters (REF 01160530,
PN 073-0035-01)
• One 18R filter (REF 01448895,
PN 073-0034-01)
• Hex wrench
Time: 15 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Clogged filters create an insufficient flow rate and produce air bubbles.
NOTE
A set of filters is included in the supplies kit. To avoid system down-time, replace the
filter with the one in the kit, resume operation, and then clean and store the removed filter
for the next scheduled maintenance.
1 Pure-water bottle
Figure 5-24. Pure-water bottle
1. Ensure that the instrument is in READY mode.
2. Remove the silicon return hose (1) from the top front of the pure water bottle.
The filter is contained in a metal filter holder attached to the end of the return hose.