122 ADVIA 2400 Operator’s Guide
Location Position Wash Solution
1
CTT-15 ISE Detergent Solution
1
CTT-16 Deionized water
1
CTT-49 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
1
CTT-50 Deionized water
1
CTT-51 Deionized water
2
RTT1-47 Reagent Probe Wash 1
2
RTT1-48 Reagent Probe Wash 2
2
RTT1-49 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
2
RTT1-50 Deionized water
3
RTT2-47 Reagent Probe Wash 1
3
RTT2-48 Reagent Probe Wash 2
3
RTT2-49 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
3
RTT2-50 Deionized water
Figure 5-14. Location of wash solutions on the CTT and RTT trays
1. At the Operation Panel, select Wash.
2. Ensure the 10-mL tube at CTT (
1) position #51 contains DI water.
TIP
By choosing CTT position #50 for WASH 2 and CTT position #51 for WASH 3, you
only need to refill the CTT container once.
3. Ensure the container at RTT1 (
2) and RTT2 (3) position #50 contains DI water.