140 ADVIA 2400 Operator’s Guide
CAUTION
Ensure that the level-sensor connector does not get wet to avoid damaging it.
7. Refill the bottle with cuvette conditioner.
8. Return the bottle to the same position on the shelf in the cabinet.
9. Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
10. Insert the filter and hose, then fasten the cap.
NOTE
Make sure that the filter holder is located at the bottom of the bottle.
11. Prime the fluid lines:
NOTE
If you are cleaning other bottles, wait to perform this step for all fluid lines.
a. At the Operation Panel, select
Prime.
b. At the Prime Settings dialog box, select
Prime 2 and type 10 for the Number of
times in all fields.
c. Select
Execute.