Siemens AG SPR2-330.815.01.03.02 ARCADIS Avantic
12.06 CS PS SP
General Information 15
Page 15 of 66
Medical Solutions
Fig. 2: Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
Pos. 1 = System
Pos. 2 = Application part type B (if available)
Pos. 3 = Measurement setup (integrated into measuring device)
If the direct measurement of the leakage current is used (measurement setup according
to (Fig.3/p.15)), the system must be insulated during the measurement and must not be
touched.
Fig. 3: Measuring circuit for direct measurement of the system leakage current in compliance