18 General information
ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG
12.06 CS PS SP
Page 18 of 64
Medical Solutions
and the measuring device used (designation and serial number).
NOTE
For evaluation purposes, the first measured value and the values
documented during previous maintenance or safety checks must
be compared to the measured values. A sudden or unexpected
increase in the measured values may indicate that a fault has
occurred in the primary power supply circuit (insulation damage,
damage from moisture, defective interference suppressor, etc.) -
even if the limit value of 2.5 mA is not exceeded.