Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
General Start-up 25
Page 25 of 64
Medical Solutions
Function check of the emergency stop switches 2.1
Connect the system to the voltage and boot it.
Click "Emergency" for the patient registration.
Operate the EMERGENCY STOP switch on the C-arm.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
Disengage the EMERGENCY STOP switch.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors