Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
Specific system settings 39
Page 39 of 64
Medical Solutions
• Switch on the system and wait for the system to boot.
NOTE
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut-
down system” button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.
Configuring the network (optional) and remote service 0
• Configure the network according to the Configuration Guide,
SPR2-310.843.01.01.xx.xx and the remote connection according to Installation, Sie-
mens Remote Service, SP00-000.816.02.01.xx.xx.
NOTE
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and
then switched on again using the power buttons on the monitor
cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut-
down system” button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.