8 General information
ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG
12.06 CS PS SP
Page 8 of 64
Medical Solutions
Safety information 1.2
General safety information (in existing documents) 0
WARNING
Danger of injuries, death or material damage.
Non-compliance can lead to death, injury or material damage.
Please note:
¹ The product-specific safety information in these instruc-
tions,
¹ The general safety information in TD00-000.860.01... and
¹ The safety information in accordance with ARTD Part 2.
General electrical safety information 0
WARNING
Electrical safety!
Non-compliance can lead to severe injury or even death, as well
as material damage.
¹ Parts under electrical voltage are accessible when the
covers are open. To avoid danger, disconnect the system
from the power supply before opening the covers. Dis-
connect the power plug.
¹ If an uninterruptible power supply (UPS) is installed in
the system, the voltage output of the UPS must also be
deenergized or the voltage output plug must be discon-
nected.
¹ If work steps must be performed using electrical power,
the general safety information according to
TD00-000.860.01... must be observed.
Main system control board D1 and power board D2
WARNING
Danger high voltage!
When the system is switched off, there is still electrical voltage on
boards D1 and D2 of the main system.
Disregarding safety precautions can result in death or serious
bodily injury.
¹ Therefore before working on one of these boards, make
sure that the LED V400 has gone out. This is the case 3
minutes after the ARCADIS Varic system is switched off.