26 General Start-up
ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG
12.06 CS PS SP
Page 26 of 64
Medical Solutions
Checking the temperature indicator 2.2
Check the temperature indicator on the outside of the image intensifier housing.
¹ If the inner, square field of the indicator is white, the temperature has not been
exceeded. Remove the temperature indicator.
¹ If the indicator is discolored (inner field black), proceed according to IQ docu-
ment RXD0-000-038.01.xx.xx.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors