36 Specific system settings
ARCADIS Varic SPR2-310.815.01.07.02 Siemens AG
12.06 CS PS SP
Page 36 of 64
Medical Solutions
• Make country-specific changes and click <Save>.
• Click <Home>.
• Perform a system shutdown.
• Switch on the system and wait for the system to boot.
NOTE
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the “Shut-
down system” button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.
• If a change is made to "Block time settings", check this function.
Country-specific adaptation of the maximum tube current 0
NOTE
A maximum tube current limit is required - as far as known - only
in Denmark.
• Start service program and log in as described under "Service Login".
"Warning time limit" "Blocking time limit"
USA
Service mask
input in 1/10 min.
5 min.
50
0 min.
0
Europe
Service mask
input in 1/10 min.
5 min.
50
10 min. (5 min. + 5 min.)
50