Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
Specific system settings 35
Page 35 of 64
Medical Solutions
Fig. 14: Main System_Configuration_Buzzer
Under Configuration click <Buzzer>.
Make changes according to the following description:
Block time settings
- Warning time limit: The buzzer sound starts when the entered radiation time has
elapsed.
- Block time limit: After the entered radiation time has elapsed - without reset of the
“warning time limit" - the radiation is blocked the next time it is selected. Every reset
of the “warning time limit” also resets the “block time limit” to “0”.
Buzzer mode (direct, cassette)
- After exposure:Buzzer sound after the exposure.
- During exposure:Buzzer sound during the exposure.
Buzzer mode (indirect, I.I.)
- Buzzer off:Buzzer sound off. Exceptions: Push mode and “warning time limit” has
ended.
- Buzzer on: Buzzer sound always during radiation.
The country-specific regulations must be observed for the max. fluoroscopy time and the
radiation blockage.
Factory setting:

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals