EasyManuals Logo

Siemens ARCADIS Varic Start-Up Instructions

Siemens ARCADIS Varic
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
Specific system settings 43
Page 43 of 64
Medical Solutions
Select “ASPIA settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
ExamSet
Select “ExamSet” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Security-Settings
For systems with an HIPAA configuration
Select “Security-Settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
MainSystem
Select “MainSystem” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Exiting the backup
Close the window via <Home>.
Remove the backup CD ROM from the CD ROM drive.
Store the CD ROM with the other system documentation.
NOTE
It is not possible to back up the database (patient images)!
Store the patient images in an archive.
Saving to CD ROM as a long-term archive is prohibited.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals