Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
Specific system settings 43
Page 43 of 64
Medical Solutions
Select “ASPIA settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
ExamSet
Select “ExamSet” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Security-Settings
For systems with an HIPAA configuration
Select “Security-Settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
MainSystem
Select “MainSystem” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Exiting the backup
Close the window via <Home>.
Remove the backup CD ROM from the CD ROM drive.
Store the CD ROM with the other system documentation.
NOTE
It is not possible to back up the database (patient images)!
Store the patient images in an archive.
Saving to CD ROM as a long-term archive is prohibited.